Hi,

 I tried to reproduce your situation, and confirmed that the DoseScorer
does not give correct answer with "/grdm/analogueMC false" and 
"/grdm/splitNuclei 10". I checked the particle
weight at the both of G4PSDoseScorer and SteppingAction and found
the particle weights are larger than expected in both case.

But, you wrote that the dose calculated at SteppingAction was 
correct. If the particle weight is wrong, the dose calculated at
SteppingAction should give wrong value, too.
So, it is very helpful if you give me your code of SteppingAction,
or if you could check the value of particle weight at your 
SteppingAction.

Best Regards,
Tsukasa Aso

Inline Depth:
 1 1
 All All
Outline Depth:
 1 1
 2 2
 All All
Add message: (add)

1 None: Re: use G4PSDoseDeposit class together with variance deduction techniques   (long - 10 Aug, 2007)
3 None: Re: use G4PSDoseDeposit class together with variance deduction techniques   (Tsukasa Aso - 20 Aug, 2007)
 Add Message Add Message
to: "Re: use G4PSDoseDeposit class together with variance deduction techniques"

 Subscribe Subscribe

This site runs SLAC HyperNews version 1.11-slac-98, derived from the original HyperNews


[ Geant 4 Home | Geant 4 HyperNews | Search | Request New Forum | Feedback ]