On Thu, 24 Jan 2019 19:56:01 GMT, Toth Arpad wrote:
>  I want to simulate a CBCT dose in Cirs IMRT thorax phantom. So I made the body
>  of phantom, than subtracted wholes in phantom to put the organs inside (lungs,
>  spine...). Since I have to rotate the phantom i made the organs as daughters of
>  phantom. The problem is that when I set lungs as scoring volume I', getting the
>  dose, but when I select the spine or the isocenter I'm getting 0 Gy. 

You don't need to, and must not, subtract holes before placing daughters. Just keep your mother "body" volume whole, and place the organ daughters in it. Geant4 does not like it when a daughter volume is placed "outside" the space occupied by its putative mother volume.

  -- Michael Kelsey