Re: Question Medical Linac Run Time Error in GEANT4.8.0.p01 (Gocha Khelashvili)
Date: 16 May, 2006
From: Michela Piergentili <Michela Piergentili>

Hi Gocha, probably you don't set the dimensions of the water phantom that is the detector, so the program takes these dimension=0.

In the last release I introduced a messenger to make the user able to change the dimension of the phantom in the macro used to run the example.

Can you try to run the example with one of the macros provided? In these macros you can find the command line: /Phantom/dimension 15. cm which set the dimension of the cube (phantom) to 15 cm

Cheers Michela