Message: Re: Spectrum 6MV Linac Not Logged In (login)
 Next-in-Thread Next-in-Thread
 Next-in-Forum Next-in-Forum

Idea Re: Spectrum 6MV Linac 

Forum: Medical Applications
Re: None Spectrum 6MV Linac (Anatoly Baulin)
Date: Mar 18, 13:46
From: Delvan Neville <Delvan Neville>

Looks like you've got a bremsstrahlung spectrum with a peak at the 511 keV electron/positron annihilation energy from pair production processes. It might look abnormal if you are comparing it to google results, since those are usually for imaging x-ray machines operating in the hundreds of kVp-max range (pair production threshold energy is 1.022 MeV).

Delvan

On Mon, 18 Mar 2019 16:16:35 GMT, Anatoly Baulin wrote:

> Hi colleagues.
> 
> 6.4 MV electrons incident on a tungsten target,
> 
> then generated photons registered by the sensitive detector.
> 
> In result, in my opinion, the energy spectrum of photons is obtained
> very strange
> 
> Can anyone give any comments?
> 
>    Attachment:
>       http://hypernews.slac.stanford.edu/HyperNews/geant4/get/AUX/2019/03/18/09.15-61243-Spectrum6MV.png
> 